Applicable Materials |
All homogeneous materials which allow transmission of Ultrasonic energy through it like most of metals, hard plastics, glass, ceramics etc. |
Measuring Range |
1mm to 300mm (Specified range depends on test material, transducer, surface and temperature.) |
Probe |
Dual crystal (Transmitter Receiver) type. |
Key Pad |
Sealed tactile type |
Velocity |
1000 m/s to 9999 m/s. |
Display |
LCD type. It displays Thickness, Velocity, Battery check, Coupling check, and stored readings with IDs. |
Resolution |
0.1 mm(0.01mm in EDISON-1P) |
Accuracy |
0.1mm (up to 60mm) {with standard 0.5% (above 60 mm) probe DT 201} |
V- path Correction |
Built-in for better accuracy. |
Calibration |
One step calibration.(for steel). |
Battery |
2 Penlite Dry Cells (AA) or NiCd. |
Operating time |
18 to 20 hours.(with alkaline batteries). |
Low Battery |
Indication by Blinking of LCD. |
Auto Power off |
3 minutes after switching on/ last reading |
Size mm |
125(H) x 65(W) x 30(D). |
Weight |
225 Grams (With batteries). |